Site map | Administration
STIMOGYN BIO 2001 Cliquer pour agrandir

STIMOGYN BIO 2001

Uro-gynaecological stimulator with a larger LCD monitor and new currents

icone pdf

BIO 2001 system fits :
  • channel 1 with EMG input or stimulation output
  • channel 2 only EMG electrodes>
Specially fitted for uro-gynaecological re-education :
  • urinary, anal incontinence
  • vaginal and sphinteral muscle re-education
  • antalgy after delivery
Operating mode :
  • stimulation only
  • dual bio-feed-back only
  • stimulation and dual bio-feed-back alternately
STIMULATION MODE

Personnal choice of current parameters :
  • wawe from : biphasic, zero mean current (symetrical, rectangular)
  • frequency : 1 to 100 Hz
  • pulse with 50 to 1000 microseconds
  • working time : 0,5 to 40 seconds
  • resting time : 0 to 40 seconds, allowing continuous stimulation
  • rising time : 0 to 5 seconds
  • frequency wobbulation
  • monophasic or biphasic current for muscles partly denerved
  • impulse width adjustable from 10 ms to 1000 ms
  • adjustement of the slope : I Max reached in adjustable time from 10 ms to 500 ms (manual adjustement)
  • adjustement of stimulation type : emission of adjustable frequency or by salves
Recorded programs
  • automatic gain calibration
  • display of DECONTRACTION and CONTRACTION voltage
  • automatic noise elimination filter allowing use without reference electrode
  • sensitivity : 50 to 100 microvolts
  • contractions display on 2 luminous scales and LCD monitor
  • loud speaker bio-feed-back control
  • RS 232 output allowing display on PC Compatibilities, printer
  • AUDIO channel on a headphone in OPTION


CHARACTERISTICS
  • 220 VAC - 50 Hz supply
  • mechanical outlines : 220 x 310 x 90 mm
  • norm IEC 610-1